systematic error

ERRORS OF MEASURING TEMPERATURE WHILE REDUCING LINEAR DIMENSIONS OF LIQUID-IN-TUBE THERMOMETERS

The methodological bases  for  the study of  the systematic error of  temperature measurement by  liquid-in-tube thermometers  and microthermometers with  diminution  of  the  capillary  diameter,  that  is, when  passing  from  thermometers  to microthermometers, in the  field of predominance of gravitational forces over surface tension  forces, are given.

Heat Exchange between Thermometer Well and Pipe Wall in Natural Gas Metering Systems

This paper deals with the accuracy of gas flow temperature measurement in natural gas metering systems. The effect of heat exchange between thermometer well and the pipe wall on the accuracy of gas flow temperature measurement is investigated. A mathematical model is proposed to calculate the additional systematic error of gas flow temperature measurement caused by heat exchange between thermometer well and the pipe wall. Calculation results were compared to the corresponding experimental values of this error.

Mathematical processing of results of double equally accurate measurements

Purpose. Run trial of criteria importance of systematic errors in the mathematical processing of results of measurements of double equally accurate and install the most efficient one. Thus, to determine how to calculate the probability of trust and develop a method of exclusion bias of the mean values of double equally accurate measurements. Methods of resolving the issues raised based on the comparative analysis formulas, experimental results of calculations, developing proposals and reports. Established that the literature significance tests are not the same bias. Results.